- Get link
- X
- Other Apps
- Get link
- X
- Other Apps
The Triaxial Compression Test is one of the most reliable ways to measure the shear strength of soil. Here I will explains how shear strength parameters of soil are determined using the triaxial compression test. It outlines the purpose of the test, the equipment involved, and how controlled loading helps measure the soil’s response under different stress conditions. This method is widely used in geotechnical engineering to evaluate stability and predict how soil will behave in the field.
Shear Strength Parameters by Triaxial Compression Test
Scope
• This test method covers determination of the
strength and stress-strain relationships of a cylindrical specimen of either
undisturbed or remolded cohesive soil. Specimens are subjected to a confining
fluid pressure in a triaxial chamber. No drainage of the specimen is permitted during the test. The specimen
is sheared in compression without
drainage at a constant rate of axial deformation
(strain controlled).
• It provides data for determining undrained strength properties and stress-strain relations for soils. This test method provides the measurement of total stresses applied to the specimen i.e. the stresses are not, corrected for pore water pressure.
Terminology
Failure: It refers
to the stress condition at failure for a test specimen. Failure
is often taken to correspond to the maximum
principal stress difference (deviator stress) attained or the principal stress
difference (deviator stress)
at 15% strain, whichever is obtained first during
the performance of a test.
Significance and Use
•
In this test method, the shear strength of a
soil is determined in terms of the total stress, therefore, the resulting strength
depends on the pressure developed
in the pore water during loading. In this method, water
flow is not permitted from or into the soil specimen as the load is applied,
therefore the resulting pore pressure, and hence strength, differs from that
developed in the case where drainage can occur.
•
If the test specimen is 100% saturated,
consolidation cannot occur during application of confining pressure nor during shearing
stage since drainage is not permitted. Therefore, if several specimens
of the same material are tested, and if they are all at approximately the same water content and void ratio
when they are tested, they will have approximately the same undrained shear
strength. The Mohr failure envelope will usually be a horizontal straight line
over the entire range of confining stresses applied to the specimens if the
specimens are fully saturated.
• If the test specimens are partially saturated or
compacted specimens, where the degree of saturation is less than 100%,
consolidation may occur when the confining pressure is applied and during
shear, even though drainage is not permitted. Therefore, if several partially
saturated specimens of the same material are tested at different confining
pressures, they will not have the same undrained shear strength, thus, the Mohr
failure envelope for UU triaxial tests on partially saturated soils is usually
covered.
• The UU triaxial strength is applicable to
situations where the loads are assumed to take place so rapidly that there is
insufficient time for the induced pore water pressure to dissipate and for consolidation to occur during
loading period (i.e.
drainage does not occur).
• Shear strength determined using the procedure
may not apply where loading
conditions in the field
differ significantly from those used during testing.
Axial loading device: The axial loading device
shall be screw jacked driven
by an electric motor through a geared transmission, a hydraulic loading
device, or any other compression device with sufficient
capacity and control to provide the rate of loading prescribed. The rate of advance of the loading
shall not deviate
by more than ±5% from the selected
value. Vibrations during application of loading shall
be sufficiently small not to cause dimensional changes
in the specimen.
NOTE: A loading
device may be said to provide sufficiently small vibrations if there are no visible ripples in a glass of water
placed on the loading platen when the device is operating at a speed at which
the test is performed.
Axial
Load Measuring Device: The axial load measuring device shall be a load
ring, electronic load cell, hydraulic load cell, or any other load measuring
device capable of measuring the axial load to an accuracy
of 1% of the axial load at failure and may be a part of the axial loading device.
Triaxial Compression Chamber: It should
consist of a top plate
and a base plate separated by a cylinder. The cylinder shall be constructed of any material
capable of withstanding the applied pressure. It is desirable to use a transparent
material or have a cylinder provided with viewing ports so that the behavior of
the specimen may be observed. The top plate shall have a vent such that air can be forced
out of the chamber as it is filled. The base plate shall have an inlet
through which the pressure fluid is supplied
to the chamber to apply cell
pressure.
Axial Load Piston: The piston passing
through the top of the chamber and its seal must be designed so that the variation in axial load due to friction does not exceed
0.1% of the axial load at
failure.
Pressure
Control Device: It shall be capable of applying and controlling the chamber
pressure to within ±2kPa (0.25psi)
for pressures less than 200kPa (28psi) and to within ±1% for pressures greater than
200kPa (28psi). This device may consist of a reservoir connected to the triaxial
chamber and partially filled with the chamber fluid
(usually water), with the upper part of the reservoir connected
to a compressed gas supply;
the gas pressure being controlled by a pressure regulator and measured
by a pressure gauge, electronic pressure transducer, or any other device capable of measuring to the
prescribed tolerance. However, a hydraulic system pressurized by deadweight
acting on a piston or any other pressure maintaining and measurement device
capable of applying and controlling the chamber pressure to a tolerance
prescribed in this section may be used.
Deformation Indicator: The vertical deformation of the specimen
shall be measured
with an accuracy of at least
0.03% of the specimen height. It shall have a range of at least 20% of the
height of the specimen, and may be a dial indicator, linear variable
differential transducer (LVDT), extensometer or other measuring
device meeting the requirements
for accuracy and range.
Rubber Membrane: It is used to incase the specimen and shall provide
reliable protection against leakage.
They shall be carefully inspected
prior to use, and if any flaws or pinholes are evident, they shall be
discarded. To offer minimum restraint to the specimen, the upstretched membrane
dia shall be between 90-95% of that of the specimen. Its thickness
shall not exceed 1% of the diameter
of the specimen. Shall be sealed to the
specimen base and cap with rubber O-rings for which the unstressed
inside dia is between 75-85% of the dia of the cap and base or by any
method that will produce a positive seal.
Sample Extruder: It shall be capable
of extruding the soil core from the sampling tube in
the same direction of travel in which the sample entered the tube and with
minimum disturbance of the sample. If the soil core is not extruded vertically,
care should be taken to avoid bending stresses on the core due to gravity.
Conditions at the time of sample removal may dictate
the direction of removal, but the principal concern is to keep the degree
of disturbance minimal.
Specimen
Size Measurement Device: Devices used to measure the height and dia of the
specimen shall be capable of measuring the desired dimension
to within 0.1% of its actual length and shall be
constructed such that their use will not disturb the specimen.
Miscellaneous
Apparatus: Specimen trimming and carving tools including a wire saw, steel
straightedge, miter box and vertical trimming lathe, apparatus for preparing
compacted specimens, remolding apparatus, water content cans and data sheets
shall be provided as required.
Test Specimens
Specimen
Size: They shall be cylindrical and have a min dia of 3.3cm (1.3”). The H/D
ratio shall be between 2-2.5. The largest particle size shall be smaller than
one-sixth the specimen dia. If, after completion of the test, it is found based
on visual observation that oversize particles are present, indicate this
information in the report of test data.
NOTE: If oversize particles are found
in the specimen after testing, a particle-size analysis may be performed
in accordance with the test method
D422 to confirm the visual observation and the results provided
with the test report.
Undisturbed Specimens: Prepare undisturbed specimens
from large undisturbed samples or from samples secured in accordance with practice
D1587 or other acceptable tube sampling procedures. Specimens obtained by the
tube sampling may be tested without trimming
except for cutting
the end surface plane and perpendicular to the longitudinal axis of the specimen, provided soil characteristics are such that no significant disturbance results
from sampling. Handle specimens carefully to minimize disturbance changes in
cross section, or change in water content. If compression or any type of
noticeable disturbance would be caused by the extrusion device, split the
sample tube length vise or cut the tube in suitable sections to facilitate
removal of the specimen with minimum disturbance. Prepare trimmed specimens, in
an environment such as controlled high humidity room where soil water content
change is minimized. Where removal of pebbles or crumbling resulting from trimming causes
voids on the surface of the specimen, carefully fill the voids
with remolded soil obtained from the trimmings. When the sample condition
permits, a vertical trimming lathe may be used to reduce the specimen to the
required dia. After obtaining the required dia, place the specimen in a miter
box and cut the specimen to the final height with a wire saw or other suitable
device. Trim the surfaces with the steel straightedge. Perform one or more water
content determinations on material trimmed
from the specimen in accordance with test method D2216.
Compacted Specimens: Soil required for compacted specimens shall be thoroughly mixed with sufficient water to produce the desired water
content. If water is added to the soil, store the material in a covered
container for at least 6hrs prior to compaction. Compacted specimens may be
prepared by compacting material in at least 6 layers using a split mold of
circular cross section having dimensions meeting the requirements enumerated in
the next section. Specimens may be compacted to the desired
density by either: o kneading or tamping each layer until the accumulative mass of the soil placed
in the mold is compacted to a known volume:
by adjusting the number of layers, the number of tamps
per layer, and the force per tamp.
The top of each layer shall be scarified prior to the addition of material
for the nest layer. The tamper used to compact the material shall have dia
equal to or less than ½ the dia of the mold. After a specimen
is formed, with the ends perpendicular to the longitudinal axis, remove the mold and determine
the mass and dimensions of the specimen.
Perform one or more water content determinations or
excess material used to prepare the specimen in accordance with D2216.
NOTE:
It is common for the unit weight of specimen after removal from the mold to
be less than the value based on the volume of the mold. This occurs because of
specimen swelling after removal of the lateral confinement due to the mold.
Procedure
•
Place the membrane on the membrane stretcher.
Place the rubber membrane around the specimen
and seal it with the cap and based with O-rings. A thin coating
silicon grease on the vertical surfaces of the cap or
base will aid in sealing the membrane.
Membrane Stretcher with suction tube, rubber
membrane, O-rings, porous disc
•
With the specimen
encased in the rubber membrane, which is sealed
to the specimen to the specimen cap and base positioned in
the chamber, assemble the triaxial chamber.
•
Place the chamber in position in the axial
loading device. Be careful to align the axial loading device, the axial
load-measuring device, and the triaxial chamber to prevent the application of a
lateral force to the piston during testing. Attach the pressure-maintaining and measurement device and fill the chamber
with the confining liquid. Adjust the pressure-
maintaining and measurement device to the desired chamber pressure and apply
the pressure to the chamber fluid. Wait approximately for 10min after the
application of the chamber pressure
to allow the specimen to stabilize under
the chamber pressure
prior to the application of the axial load.
NOTE: In some cases the chamber will be filled
and the chamber
pressure applied before placement in axial loading device.
NOTE: Make sure the piston
is locked or held in place by the axial loading device before
applying the chamber pressure.
NOTE: The waiting
period may need to be increased for soft or partially saturated soils
•
Apply the axial load to produce
axial strain at a rate of 1%/min for plastic materials and 0.3%/min for brittle
materials that achieve max deviator stress at approximately 3-6% strain.
Continue the loading to 15% axial strain, except loading may be stopped when
the deviator stress has peaked then dropped 20% or the axial strain has reached
5% beyond the strain at which the peak in deviator
stress occurred.
•
Record load and deformation values to 3
significant digits at about 0.1, 0.2, 0.3, 0.4 and 0.5% strain; then at
increments of about 0.5% strain to 3%; and, thereafter at every 1%. Take
sufficient readings to define the stress-strain curve; hence, more frequent
readings may be required in the early stages of the test and as failure is
approached.
•
After completion of the test, remove the test
specimen from the chamber. Determine the water content of the test specimen
using the entire specimen, if possible.
•
Prior to placing
the specimen (or its portion)
in the oven to dry, take a snap of the specimen showing the mode of failure
(shear plane, bulging etc.).
Calculations
•
Measurements and calculations shall contain 3 significant figures.
•
Calculate the axial strain for a given applied axial load, as follows:
Ԑ = ∆H/Ho
Where: ∆H = change in height of the specimen
as read from the deformation indicator
Ho = Initial height of the specimen minus any change in length prior to loading
• Calculate average cross-sectional area, A,
for a given applied axial load as follows.
A = Ao / (1 - Ԑ)
Where: Ao = Initial average cross-sectional area of the specim.
Calculate the principal stress difference (deviator stress), σ1 – σ3 for a given applied axial load as follows:
σ1 – σ3 = P/A Where:
P = measured
applied axial load
Calculate the major and minor principal total stresses at failure as follows:
σ3
=
major principal stress = chamber pressure σ1 = major principal
total stress=deviator stress at failure
plus chamber pressure
= ∆p+ σ3.
Calculate initial degree of saturation of the specimen using initial mass and dimensions.
NOTE: The
Gs determined in
accordance with D854 is required for calculation of the saturation. An assumed
Gs may be used provided
it is noted in the test report.
•
Plot Mohr’s circle,
failure envelope and determine C and ϕ.
Conclusion
Comments
Post a Comment